The response times during anaesthesia of pulse oximeters measuring oxygen saturations during hypoxaemic events.

Broome, I J; Harris, R W; Reilly, C S · Anaesthesia · 1992

cross_sectional · Level IV

Where this comes from

Abstract

Episodes of desaturation were recorded simultaneously by computer from two Ohmeda Biox 3700 pulse oximeters, one with an ear and one with a finger probe, on patients undergoing anaesthesia. Over a period of 6 months, 28 episodes of desaturation were detected. Analysis of the recordings showed the mean minimum saturations recorded for ear and finger probes were 86.3% and 83.5% respectively (p less than 0.01). The mean delay for finger compared to ear pulse oximetry was 4.4 s (p less than 0.01). Analysis at different saturation levels showed finger probe responses to be significantly slower than ear probe responses at saturations equal to and above 91% (p less than 0.05). At saturation levels of 90% or less no significant difference in probe response times were found.