To determine the reliability of a recently developed electrochemical meter to rapidly (within 60 seconds) measure blood alcohol concentration (BAC) in the emergency department. A prospective study comparing the meter data with that of immunoassay and gas chromatography criterion standards undertaken during a ten-week period. Adult ED of a municipal hospital. Three hundred eighty-three consecutive patients with altered mental status or suspected alcohol intoxication. Each patient underwent routine phlebotomy, and blood samples were obtained for meter and immunoassay BAC determinations. The first 60% of patients also underwent gas chromatography BAC determination. Two hundred nineteen patients (60%) had BAC detectable by both meter and immunoassay. BAC measurement by the meter correlated strongly with immunoassay and gas chromatography determinations (Pearson's correlation coefficient, r = .94; P less than .00000001 for both correlations). The electrochemical meter provides a rapid and reliable BAC measurement in the ED.
Medical subject headings
- Alcoholic Intoxication
- Electrochemistry