Functional magnetic ventilation in dogs.

Lin, Vernon W; Hsiao, Ian; Deng, Xiaoming; Lee, Yu-Sheng; Sasse, Scott · Arch Phys Med Rehabil · 2004

basic_science · Level V

Where this comes from

Abstract

To investigate the efficacy of the magnetic stimulation of inspiratory muscles as an alternative to mechanical ventilation and functional electric stimulation. A prospective before-after trial. Functional magnetic stimulation laboratory in a Veterans Administration health care system. Six male mongrel dogs, each weighing between 25 and 35 kg. Commercially available magnetic stimulators with a round magnetic coil were used. The center of the magnetic coil was placed posteriorly over the C5-7 vertebrae of the spinal cord transected dogs. Magnetic stimulation parameters were set at 80% intensity, 20 Hz, and a 1.2-second on and 3.8-second off pulse train. The major outcomes were changes in tidal volume (VT), tracheal pressure (Ptr), and arterial partial pressure of carbon dioxide (PaCO2) and oxygen sustained by magnetic stimulation over time. The average Vt and Ptr produced during functional magnetic ventilation (FMV) were.47+/-.07 L and -4.7+/-.51 cmH2O, respectively. Blood gas data showed that PaCO2 increased from a baseline of 33 to 75 mmHg, whereas pH decreased from 7.33 to 6.99 at the end of the 1-hour FMV period. FMV was achieved for 2 hours in dogs with C2 spinal cord transection. Additional refinements in magnetic stimulation are needed to improve ventilation in animals.