Post-dural puncture headache in young orthopaedic in-patients: comparison of a 0.33 mm (29-gauge) Quincke-type with a 0.7 mm (22-gauge) Whitacre spinal needle in 200 patients.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 1539481.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The incidence of postdural puncture headache (PDPH) following spinal anaesthesia with a 0.33 mm (29-gauge) and two types of 0.7 mm (22-gauge) Whitacre needles was investigated in 400 patients less than 40 years old. The incidence of PDPH was 2% in the 0.33 mm group, and 3.5% in the 0.7 mm group. Headache of other origin was seen in 12 patients (6%) in the 0.33 mm and in five patients (2.5%) in the 0.7 mm group. These differences were not significant. The failure rate was significantly higher in the 0.33 mm group (8.5% vs 2%) than in the 0.7 mm group (P less than 0.05). It is concluded that the 0.33 mm needle is associated with a low incidence of PDPH in young patients, but has a significantly higher failure rate than the Whitacre 0.7 mm needle, which is also a suitable choice in this age-group because of its ease of handling and the low incidence of PDPH.
Medical subject headings
- Anesthesia, Spinal
- Headache
- Needles
- Spinal Puncture