Three- to six-year results with the Ultima metal-on-metal hip articulation for primary total hip arthroplasty.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 15457418.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
One hundred seventy-one primary total hip arthroplasties were evaluated in a prospective, randomized study. Ninety-five involved a metal-backed cup with an all-metal liner and 76 involved a metal-backed polyethylene cup that was used as the control. All were implanted with an S-ROM cementless femoral component with a 28 mm head. The mean follow-up period was 3.7 years (range, 3.0-5.7). The average postoperative Harris hip score was 95.4 (range, 65-100) for the metal-on-metal group and 96.1 (range, 65-100) for the metal-on-polyethylene group. Radiographic results were not statistically different between the 2 groups. Early results show the metal-on-metal articulation has been successful to date and justify continued clinical use.
Medical subject headings
- Polyethylene
Anatomy
- hip