Secondary knee instability caused by fracture of the stabilizing insert in a dual-articular total knee.

Boesen, Morten P; Jensen, Tim Toftgaard; Husted, Henrik · J Arthroplasty · 2004

case_report · Level V

Where this comes from

Abstract

A case of a fractured polyethylene stabilizing insert causing secondary knee instability in a Dual-articular total knee arthroplasty (TKA) is presented. A 65-year-old woman who underwent surgery with a Dual-articular TKA 4 years earlier had a well-functioning prosthesis until a fall, after which she began complaining of an intermittent, audible clicking; pain; and instability. At surgery, a fractured polyethylene peg was discovered, and the tibial insert was replaced. At follow-up, she remains asymptomatic. This is the first description of a fractured stabilizing insert in a Dual-articular knee.