The comparability of estimates of retroilluminated lens opacities as judged from film-based and digital imaging.

Klein, Barbara E K; Klein, Ronald; Hall, Ellen R; Lee, Kristine E; Jensen, Karl · Am J Ophthalmol · 2004

cross_sectional · Level IV

Where this comes from

Abstract

Neitz film-based retroillumination cameras, the standard for documenting retroilluminated lens opacities for epidemiologic studies, are no longer produced. A digital imaging system is now available for imaging these opacities. We sought to compare gradings of images from both systems. Comparison of technique. One hundred fourteen lenses were imaged with both systems and graded according to protocols. Concordance between the methods was compared using kappa statistics. There was moderate concordance for cortical opacities (kappa = 0.63) and good concordance for posterior subcapsular opacities (kappa = 0.83). Grades from digital images slightly underestimated the frequency and severity of cortical cataract. Digital imaging of retroilluminated lens opacities results in similar classification of the severity of opacities. It will be useful for epidemiologic studies of cortical (CC) and posterior subcapsular cataracts (PSC).

Medical subject headings