A new mechanical testing device for measuring anteroposterior knee laxity.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 15494340.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The authors assessed a new instrument, the Rolimeter, for the measurement of anteroposterior translation in the knee; it was compared to the KT-1000 arthrometer. To determine if the Rolimeter offers a valid method for the measurement of anteroposterior translation that is as reproducible and reliable as the KT-1000 arthrometer. Two of 3 observers examined 16 normal subjects (32 knees) and 36 patients (72 knees) with ligament ruptures twice, using both the Rolimeter and the KT-1000 arthrometers, 30 minutes apart. Total anteroposterior translation (manual maximal Lachman test) was recorded at 20 degrees and 80 degrees of knee flexion. On average, the Rolimeter measured approximately 1 mm less anteroposterior displacement than the KT-1000 arthrometer at manual maximum stress. Rolimeter measurements were more consistent than the KT-1000 measurements as measured by our observers. Specificity and sensitivity were equivalent between the Rolimeter and the KT-1000 arthrometer. The Rolimeter is as reproducible and reliable as the KT-1000 arthrometer. It offers a valid method for the measurement of anteroposterior translation in the knee. Higher accuracy was obtained at 20 degrees of flexion for the KT-1000 arthrometer and at 80 degrees of flexion for the Rolimeter.
Medical subject headings
- Anterior Cruciate Ligament Injuries
- Joint Instability
- Knee Injuries
- Knee Joint
- Posterior Cruciate Ligament
Anatomy
- knee