Zirconia to Ti-6Al-4V braze joint for implantable biomedical device.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 15497165.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A strong, hermetic, reliable, and biocompatible ceramic-to-metal seal is essential for many implantable medical devices. Yttria-stabilized tetragonal zirconia polycrystals (Y-TZPs) and a titanium alloy Ti-6Al-4V were selected as the ceramic and metal components of the seal because both materials have excellent mechanical properties and favorable biocompatibility. A brazing method using titanium nickel (TiNi)-clad braze filler material is presented to bond the components together forming a seal. Laboratory tests show that the ceramic-to-metal seal is hermetic, strong, and resistant to electrochemical corrosion. Twenty-eight microstimulators utilizing the ceramic-to-metal seals were implanted in seven sheep to stimulate the hypoglossal nerve. When the tissue was evaluated by gross inspection at necropsy and examined histologically by a pathologist, there were no signs of local hemorrhage, infection, or hypoglossal nerve tissue damage.
Medical subject headings
- Dental Porcelain
- Electric Stimulation
- Implants, Experimental
- Materials Testing
- Titanium
- Yttrium