An American experience with metal-on-metal total hip arthroplasties: a 7-year follow-up study.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 15578549.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study reviews the clinical performance of 161 hip arthroplasties (154 patients) with the Metasul metal-on-metal articulation and an uncemented modular acetabular component. Between 1995 and 2002 clinical evaluation and radiographic follow-up of patients included Harris hip scores, patient self-assessment, and radiographs. Twelve operative site complications (7.5%) included 6 revision operations, (3.7%) and 3 other complications (1.9%) not needing reoperation. Six revision operations (3.7%) included 1 femoral revision for aseptic loosening (0.06%) and 5 acetabular revisions (3.1%). One cup revision was for recurrent dislocation, 1 for disassociation of the acetabular insert from the cup, 1 for infection, and 2 for unexplained pain. Histologic evidence did not support the diagnosis of metal hypersensitivity in either case of unexplained pain, and 1 had relief following spine surgery. A focal radiolucency, identified as calcar resorption, was observed in 9 patients.
Medical subject headings
- Hip Prosthesis
Anatomy
- hip