We investigated whether cup inclination influences serum levels of cobalt and chromium after metal-on-metal total hip arthroplasty. Out of a consecutive series, 3 groups of 20 patients each were established according to their cup inclination: greatest inclination (55 degrees -63 degrees, mean 58 degrees), intermediate inclination (44 degrees -46 degrees, mean 45 degrees), and smallest inclination (23 degrees -37 degrees, mean 33 degrees). We did not detect any statistically significant difference of serum cobalt (P = .23) or serum chromium (P = .13) levels between the 3 groups applying the Kruskal Wallis test. However, 3 patients with cup inclinations of 58 degrees , 63 degrees , and 61 degrees exhibited 9.8-53.6-fold elevated cobalt and 9.5-30.5-fold elevated chromium levels when compared with the median concentrations of this trial. We therefore recommend accurate cup placement also for metal-on-metal articulations.