Comparison of the anchoring strengths of the Kopans and Hawkins II needle-hook-wire systems.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 1561371.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine whether needle-hook-wire anchoring strength is a factor in accidental repositioning of a hook wire, 5-cm Kopans and Hawkins II hook wires were placed axially (n = 9) and coronally (n = 5) in mastectomy specimens. The tension required to pull the hook wires 1, 2, 3, and 5 cm and to increase the barb angle to 90 degrees and 180 degrees were compared. The average anchoring strength of the Hawkins II hook wire was greater than that of the Kopans hook wire, by 2.6 times (P less than .001; all distances, both angles) with the wires in the axial plane and by 1.6 times (P less than .06; intervals and angles combined) in the coronal plane.