Clinical comparison of three different anaesthetic depth monitors during cardiopulmonary bypass.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 15644019.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The lack of a gold standard complicates the evaluation and comparison of anaesthetic depth monitors. This randomised study compares three different depth-of-anaesthesia monitors during cardiopulmonary bypass (CPB) at 34 degrees C with fentanyl/propofol anaesthesia adjusted clinically and blinded to the monitors. Coronary artery bypass grafting patients (n = 21) were randomly assigned to all three possible paired combinations of three monitors: Bispectral Index (Aspect Medical), AAI auditory evoked potential (Danmeter), Entropy (Datex-Ohmeda). Indices were manually recorded every 5 min during CPB. Agreement between paired indices was classified as good, non-, or disagreement. Anaesthesia was classed as adequate, inadequate, or excessive according to recommended index values. Of the 255 paired indices recorded, 62% showed good agreement, 33% showed non-agreement, and 5% showed disagreement. Using good agreement between two monitors as a gold standard, a quarter of the measurements indicate inappropriate anaesthetic depth monitoring during CPB with clinically titrated anaesthetic depth.
Medical subject headings
- Anesthetics, General
- Cardiopulmonary Bypass
- Consciousness
- Monitoring, Intraoperative