A multicentre evaluation of the Medtronic AVE Flexible Iliac Bridge Stent in the iliac arteries (the first study).
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 15649717.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To investigate the short term and 6 month outcomes of a recently introduced iliac stent (the Medtronic AVE Flexible Iliac Bridge Stent). One hundred and sixteen patients, 122 limbs from 6 European sites with atherosclerotic occlusive disease were included if they either had a full occlusion or there was a residual gradient >10 mmHg following PTA. Clinical, haemodynamic and Duplex outcomes were recorded to 6 months. After stent placement there was no residual stenosis >30 and 8.1% of segments had a residual resting gradient of >10 mmHg. There were three local complications and three deaths at 30 days. Primary patency at 30 days and 6 months was 94.1 and 82.7%. Marked clinical improvement occurred in 87.6 and 86.2% at 30 days and 6 months. The Medtronic AVE iliac stent has good 30 day and 6 months outcomes supporting the effective use of this device in occlusive disease.
Medical subject headings
- Arterial Occlusive Diseases
- Iliac Artery
- Outcome Assessment, Health Care
- Stents