Electrocardiographic motion artifact versus electrode impedance.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 15651575.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We degraded electrocardiographic electrodes by exposing them to air for four days and evaluated them on 12 subjects. After application, we recorded the electrocardiogram (including motion artifact), missed QRS detections and electrode impedance during 5 min of arm and body movements. Missed QRS detections increased with electrode impedance but correlation was poor. Increased electrode impedance was not a reliable predictor of a poor electrode and the need to replace it.