Current of injury predicts adequate active lead fixation in permanent pacemaker/defibrillation leads.

Saxonhouse, Sherry J; Conti, Jamie B; Curtis, Anne B · J Am Coll Cardiol · 2005

prospective_cohort · Level II

Where this comes from

Abstract

The aim of this study was to determine whether current of injury can guide adequate placement of active-fixation pacing leads. Active-fixation leads cause injury to the myocardium at the time of fixation, manifested as a current of injury (COI) that may result in acute elevation of pacing thresholds. The relationship of COI to subsequent improvement in pacing thresholds is not clear. Sixty-five patients undergoing active-fixation lead implantation were enrolled. Current of injury was characterized as the duration of the intracardiac electrogram (EGM) and the magnitude of ST-segment elevation. Pacing parameters were measured up to 10 min after fixation. A total of 96 active-fixation leads were studied, and 76 leads had a current of injury. From baseline to the time of fixation, the duration of the intracardiac EGM in ventricular leads increased from 150 +/- 31 ms to 200 +/- 25 ms (p < 0.001), and the ST-segment increased from 1.5 +/- 0.2 mV to 10.0 +/- 2.0 mV (p < 0.001), with subsequent improvement in pacing thresholds from 1.5 +/- 0.4 V to 0.8 +/- 0.3 V (p < 0.001) at 10 min. Atrial leads with a current of injury had similar findings. Of the 20 leads without a COI, 5 dislodged acutely and 15 had high pacing thresholds at 10 min, requiring repositioning. The development of a COI indicates that within 10 min of fixation, pacing threshold will return to an acceptable range even if the initial measurement is high. Conversely, without a COI, lead fixation is not adequate and the lead should be repositioned.

Medical subject headings