Measurements and treatment planning calculations of electron dose distributions below bolus edges.

Nygaard, Kirsten; Odland, Odd Harald; Kvinnsland, Yngve; Nygaard, Britt; Heggdal, Jan; Muren, Ludvig Paul · Radiother Oncol · 2005

basic_science · Level V

Where this comes from

Abstract

Electron dose distributions below bolus edges of various thicknesses and angles were measured using a diode in a water phantom. The measurements were compared with treatment planning calculations using the gamma method. Below 5 mm boluses, the dose variations were acceptable for all angles, while for boluses thicker than 5mm, the lowest edge gave the smallest dose variations. The calculated dose distributions agreed well with the measured dose distributions.