Home use of the GlucoWatch G2 biographer in children with diabetes.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 15741369.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate usability, accuracy, and hypoglycemia detection of the GlucoWatch G2 Biographer (GW2B) in children aged 1 to 17 years. After a 15-hour study of device accuracy, 46 children (15 <7 years, 31 > or =7 years) with type 1 diabetes were enrolled for an extended-wear outcome study: 2 daytime and 2 nighttime 15-hr wear periods each week and blood glucose monitoring 4 times daily for 3 months. A total of 531 paired GW2B/meter readings were available for accuracy assessment. The correlation coefficients were 0.58 and 0.74 (ages <7 and > or =7 years, respectively). There was no significant change in hemoglobin A1C)or weight-adjusted insulin dose at 3 months after biographer use. Forty-two episodes of hypoglycemia were detected by the GW2B, 33 of which were confirmed by blood glucose meters. Sensitivity and specificity of audible low-glucose alerts were approximately 79% and 83%, respectively. No significant side effects were reported. The GW2B is usable and safe in children who are <7 years or older in the home setting. The GW2B can detect asymptomatic nocturnal hypoglycemia in younger children.
Medical subject headings
- Blood Glucose Self-Monitoring
- Diabetes Mellitus, Type 1
- Hypoglycemia