Beam modifying devices in the treatment of early breast cancer: 3-D stepped compensating technique.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 1574598.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A 3-D compensating technique utilizing stepped brass filters to ensure an adequate dose distribution in the breast after conservative surgery is described. No compensation for lung inhomogeneity is applied. These compensators are part of a procedure which also involves the use of a rotatable half-beam block (HBB) and of an individualized immobilization cast. Results of patient dosimetry show a good agreement with dose prescription in the target volume (measured dose/prescribed dose = 100.9% +/- 1.8% at the reference point). Phantom dosimetry is discussed.
Medical subject headings
- Breast Neoplasms
- Mastectomy, Segmental
- Radiotherapy