Usefulness of unipolar epicardial ventricular electrogram for triggering of intraaortic balloon counterpulsation during off-pump coronary artery bypass surgery in patients with hemodynamic instability complicating acute coronary syndromes.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 15781501.
- Also identified by DOI 10.1213/01.ANE.0000146941.05215.CC.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
During off-pump coronary artery bypass graft (CABG) surgery for patients requiring preoperative intraaortic balloon counterpulsation (IABP) support, errors in timing of IABP during mobilization of the heart can induce unstable hemodynamic conditions. We applied a simple technique for triggering IABP accurately during off-pump CABG: one end of an epicardial pacemaker wire is placed on the surface of the left ventricle, and the other end is linked to one of the precordial V leads. This technique provided an exact detection of R-wave potentials in any position of the heart during the off-pump CABG procedure. Ten patients with hemodynamic instability complicating acute coronary syndromes underwent off-pump CABG with this technique, including grafting to the circumflex system, safely and successfully.
Medical subject headings
- Angioplasty, Balloon, Coronary
- Coronary Artery Bypass, Off-Pump
- Electrocardiography
- Intraoperative Complications
- Pericardium