A biomechanical comparison of four reconstruction techniques for the medial collateral ligament-deficient elbow.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 15789016.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The initial strength of the intact medial collateral ligament (MCL) of the elbow and the strength of 4 reconstruction techniques were compared. Twenty cadaveric upper extremities were mounted in a custom jig with the elbow at 90 degrees , and a pneumatic cyclic valgus loading protocol was used. The mean peak load to failure was 142.5+/-39.4 N for the intact ligaments and 53.0+/-9.5 N for the docking reconstructions, 52.5+/-10.4 N for the EndoButton reconstructions, 41.0+/-16.0 N for the interference screw reconstructions, and 33.3+/-7.1 N for the figure-eight reconstructions. The peak load to failure of the MCL reconstructions was inferior compared with the intact ligament (P<.001). No difference in strength was found between the docking and single-strand medial collateral reconstruction with the use of an EndoButton for ulnar fixation (P>.05, beta=.14). Both of these reconstruction methods were stronger than the interference screw or figure-eight technique (P<.004). The optimal fixation method for a single-strand MCL reconstruction may require improved interference screws or a modified EndoButton procedure.