Performance of orientation sensors for use with a functional electrical stimulation mobility system.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 15797599.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to verify the performance of recently developed body-worn sensor packs against 3D motion analysis of trunk and lower-limb movements. Five sensor packs, each consisting of rate gyroscope and two 2-D accelerometers controlled by a microprocessor were attached to the trunk, thighs, and shanks of an able bodied subject. A 6-camera motion analysis system (MAS) recorded multiple trials of sit-to-stand movements and normal walking. Time domain signals from each sensor pack were significantly correlated (r = 0.90-0.99;p < 0.05) with a root mean square errors of less than 5 degrees when compared against the same limb angle measurements calculated by the MAS. These data demonstrate that these external sensor packs are accurate devices for measuring trunk and lower-limb sagittal plane orientation in real-time.