Precoated femoral component with proximal and distal centralizers: results at 5 to 12 years.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 15809947.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
One study, confounded by the use of crystalline polyethylene and ceramic heads, reported a high rate of early failure of a precoated femoral component with proximal and distal centralizers. The present study reports the prospective clinical and radiographic results, and 10-year survival data of 166 consecutive hybrid total hip arthroplasties using this femoral component. An A or B cement grade was obtained in 93% of hips. At a mean follow-up time of 7 years (range, 5-12 years), there were only 4 (2.4%) femoral failures. Ten-year survival of this component was 95% (confidence interval, 94%-99%). The rate of radiographic failure and revision of this component implanted with conventional polyethylene and cobalt chrome heads is similar to that reported with other "modern" femoral components.
Medical subject headings
- Arthritis, Rheumatoid
- Ceramics
- Chromium Alloys
- Coated Materials, Biocompatible
- Equipment Failure Analysis
- Femur Head Necrosis
- Hip Prosthesis
- Osteoarthritis, Hip
- Polyethylene
Anatomy
- femur
- hip