Stress risers between two ipsilateral intramedullary stems: a finite-element and biomechanical analysis.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 15809959.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Periprosthetic fracture after total joint replacement predominantly occurs at the stem tip. In this study, the effects of gap size, stem stability and cortical thickness between two press-fit, ipsilateral intramedullary stems on the tensile stresses created in the femur were investigated using finite-element models. The findings were confirmed with strain-gauge tests using a composite Sawbone femur. Gap size did not affect the level of stress on the femur. Cortical thickness had an important effect on stress distribution: peak stresses increased as bone cortical thickness decreased. Irrespective of gap size, the tips of loose stems acted as stress risers particularly with thinner cortices; the tips of well-fixed stems, however, did not.
Medical subject headings
- Equipment Failure Analysis
- Femoral Fractures
- Finite Element Analysis
- Fractures, Stress
- Hip Prosthesis
- Postoperative Complications
Anatomy
- femur
- hip