Acute locking in revision total knee arthroplasty due to disengagement of the locking screw.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 15824933.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Acute locking of the joint in a replaced knee joint is very rare. This report describes an acute locking episode of a revised modular total knee replacement, occurring more than 2 years after surgery. A disengaged screw from the modular femoral component had lodged in the joint at the inferior pole of the patella and required urgent arthroscopic removal. There was no subsequent failure of the stem-condylar junction, nor loosening of the femoral component.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Prosthesis
- Prosthesis Failure
Anatomy
- knee