Peroneal nerve damage associated with the proximal locking screws of the AIM tibial nail.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 15826626.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report a case of division of the deep peroneal nerve resulting from a drill used in the insertion of an oblique proximal locking screw in an AIM tibial intramedullary nail (DePuy). Operative findings and anatomical study indicate there is a risk of damage to the peroneal nerve associated with the oblique proximal locking screws used in this nail design. If a patient has peroneal nerve palsy after nailing of the tibia, the possibility of nerve division should be considered, so that early exploration and repair of the nerve can be performed.
Medical subject headings
- Bone Screws
- Fracture Fixation, Intramedullary
- Peroneal Nerve
- Tibial Fractures