Influence of the clearance on in-vitro tribology of large diameter metal-on-metal articulations pertaining to resurfacing hip implants.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 15833451.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Large-diameter metal-on-metal articulations may provide an opportunity for wear reduction in total hip implants because earlier studies have shown that the formation of a fluid film that completely separates the bearing surfaces is theoretically possible. In such a lubrication mode and under ideal conditions, there is theoretically no amount of wear. Studies have suggested that the two primary parameters controlling the lubrication mode are the diameter and the clearance of the articulation. The goal of the present study was to experimentally investigate the influence of these two parameters on the wear behavior of large-diameter metal-on-metal articulations pertaining to resurfacing hip implants. The results of this in vitro investigation showed that longer running-in periods and higher amounts of running-in wear were associated with larger clearances.
Medical subject headings
- Equipment Design
- Hip Prosthesis
- Metals