Tying a loop-to-strand suture: is it safe?
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 15846186.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study was undertaken to evaluate the integrity of the loop-to-strand knot when tied with square and nonidentical sliding knots. The synthetic absorbable monofilament suture poliglecaprone 25 in 0 and 2-0 suture gauges was used in this experiment. For each suture gauge, 3 groups of knots were tested: (1) single strand-to-single strand, flat square knot, (2) loop-to-single strand, flat square knot, and (3) loop-to-single strand, nonidentical sliding knot. All knots were tied with 6 throws. The proportion of knots becoming untied was compared among the 3 groups for each suture gauge. Ultimate load required to untie or break knots within each group was also evaluated. The loop-to-strand knot performed well in both suture gauges tested as long as it was tied with a flat square knot. The loop-to-strand knot tied with a nonidentical sliding knot had an unacceptably high failure rate. The loop-to-strand termination of a continuous suture may be acceptable when tied with a 6-throw flat square knot but not acceptable if tied with sliding knots.
Medical subject headings
- Dioxanes
- Polyesters
- Suture Techniques