The relationship between shelf life and in vivo wear for polyethylene acetabular liners.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 15902855.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Abstract: Although laboratory studies have correlated shelf life for gamma-irradiated-in-air polyethylene with material degradation, it is unknown whether this clinically affects wear. Therefore, we examined the relationship between shelf life and clinical wear for 152 gamma-irradiated-in-air acetabular liners. True wear rates were calculated from computer-assisted head penetration analyses of serial radiographs. Follow-up averaged 6.6 years. Shelf life averaged 11.8 months with 96% of liners having shelf lives of less than 3 years. Data analysis revealed no correlation between shelf life and true wear rates (r2 = 0.002; P = .60). This study provides useful information for surgeons concerned about shelf life as a potential cause of high wear in previously implanted components. For conventional liners in inventory less than 3 years, we found no evidence of increased wear in vivo.
Medical subject headings
- Hip Prosthesis
- Polyethylene
Anatomy
- hip