Fracture of the femur in revision hip arthroplasty with a fully porous-coated component.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 15902863.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Fifty-four consecutive femoral component revisions with uncemented fully porous-coated stems were retrospectively reviewed for the prevalence of intraoperative fracture. Fracture of the femoral diaphysis occurred during insertion of a curved stem in 8 (15%) hips. One additional distal fracture of the femur occurred during exposure of the hip. All fractures occurred with a curved femoral stem. Treatment for intraoperative fractures included open reduction and internal fixation with cortical strut grafts and cables followed by protected weight bearing. This was successful in 6 hips. Even with the use of a curved stem, we report a high prevalence of intraoperative fracture of the femur in revision hip arthroplasty.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Femoral Fractures
- Hip Prosthesis
- Intraoperative Complications
- Reoperation
Anatomy
- femur
- hip