New suction guide needle designed to reduce the incidence of biopsy-related pneumothorax: experimental evaluation in canine model.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 15914484.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In an attempt to remove air that enters the pleural space during computed tomography (CT)-guided coaxial transthoracic needle biopsy, the authors fashioned an 18-gauge experimental suction guide needle and evaluated the incidence of pneumothorax with this needle in comparison to the incidence of pneumothorax with a standard 18-gauge guide needle in a canine model. This experiment had animal care and use committee approval. Ten dogs underwent a biopsy of each lung, for a total of 20 lung biopsies. Half of the biopsies were performed by using the experimental needle (five right lungs, five left lungs), and half were performed by using a standard guide needle. CT revealed pneumothorax during the procedure and was performed to reveal pneumothorax 1 and 3 hours after the procedure. A significant reduction (P < .016) in intraprocedural lung biopsy-associated pneumothorax was found when the experimental guide needle was used.
Medical subject headings
- Biopsy, Needle
- Lung
- Needles
- Pneumothorax
- Tomography, X-Ray Computed