Comparing step-and-shoot IMRT with dynamic helical tomotherapy IMRT plans for head-and-neck cancer.
other
Where this comes from
- Record sourced from PubMed, PMID 15925457.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The goal of this planning study was to compare step-and-shoot intensity-modulated radiotherapy (IMRT) plans with helical dynamic IMRT plans for oropharynx patients on the basis of dose distribution. Five patients with oropharynx cancer had been previously treated by step-and-shoot IMRT at the University Medical Centre Utrecht, The Netherlands, applying five fields and approximately 60-90 segments. Inverse planning was carried out using Plato, version 2.6.2. For each patient, an inverse IMRT plan was also made using Tomotherapy Hi-Art System, version 2.0, and using the same targets and optimization goals. Statistical analysis was performed by a paired t test. All tomotherapy plans compared favorably with the step-and-shoot plans regarding sparing of the organs at risk and keeping an equivalent target dose homogeneity. Tomotherapy plans in particular realized sharper dose gradients compared with the step-and-shoot plans. The mean dose to all parotid glands (n = 10) decreased on average 6.5 Gy (range, -4 to 14; p = 0.002). The theoretical reduction in normal tissue complication probabilities in favor of the tomotherapy plans depended on the parotid normal tissue complication probability model used (range, -3% to 32%). Helical tomotherapy IMRT plans realized sharper dose gradients compared with the clinically applied step-and shoot plans. They are expected to be able to reduce the parotid normal tissue complication probability further, keeping a similar target dose homogeneity.
Medical subject headings
- Head and Neck Neoplasms
- Radiotherapy Planning, Computer-Assisted
- Radiotherapy, Conformal
- Tomography, Spiral Computed