Profiles in patient safety: misplaced femoral line guidewire and multiple failures to detect the foreign body on chest radiography.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 15995100.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This article uses a case report and discussion to demonstrate the concept of active and latent failures, and the "systems approach" to the reduction of adverse events in medicine. The case involves an inadvertently misplaced and retained guidewire during femoral vein catheterization using the Seldinger technique, and the subsequent failure to identify the guidewire in the chest despite several chest radiographs and a computed tomography (CT) scan read by radiologists, emergency physicians, and intensivists. This event reveals active failures in the performance of the Seldinger technique, latent failures in the design of the catheter kit, and problems with the current system of interpretation of radiographs. The authors conclude that the design of the catheter kit and the Seldinger technique should be critically examined from a human factors standpoint and that radiographic interpretation is still heavily subject to human error.
Medical subject headings
- Catheterization, Central Venous
- Diagnostic Errors
- Foreign Bodies
- Vena Cava, Inferior