Molecular magnetic resonance imaging of atrial clots in a swine model.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 16009790.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The detection and differentiation of intracardiac masses is still challenging and may include neoplasms and thrombi. The aim of this study was the investigation of a targeted, fibrin-specific contrast agent (EP-2104R) for molecular targeted magnetic resonance imaging (MRI) of left atrial clots. Chronic human thrombi were surgically implanted in the left atrial appendage of 5 swine. Molecular MRI was performed with a navigator-gated, free-breathing, cardiac-triggered 3D inversion-recovery, black-blood, gradient-echo sequence before and after systemic administration of 4 micromol/kg EP-2104R. MR images were analyzed by 2 investigators, and the contrast-to-noise ratio was calculated. Location of clots was confirmed by autopsy, and the gadolinium concentration in the clots was assessed. Before contrast agent administration, thrombi were not visible on black-blood MR images. After contrast administration, all atrial clots (n=5) were selectively visualized as white spots with a high contrast-to-noise ratio (clot/blood, 29.7+/-8.0). The gadolinium concentration in the clots averaged 74+/-45 micromol/L. The fibrin-specific MR contrast agent EP-2104R allows for selective and high-contrast visualization of left atrial clots by means of molecular targeted MRI.
Medical subject headings
- Heart Atria
- Heart Diseases