The second-generation aortic connector: six months' angiographic follow-up.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 16039230.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Recently, new mechanical anastomotic devices have been developed. Initial results appear to be equivalent to those obtained with suture. The aim of the study was to evaluate the 6-month angiographic patency and clinical results with the St. Jude Medical second-generation aortic connector for proximal aortosaphenous graft anastomosis. From September 2002 to June 2003, 45 connectors were implanted in 39 patients. Thirty-three patients with 36 connectors underwent 6-month angiographic and clinical follow-up. One connector had an early occlusion and 2 connectors and 1 vein graft were occluded at 6-month angiography, for a patency rate of 88.9% (32 of 36). No device-related complications were detected at 6-month follow-up. The second-generation aortic connector is safe and easy to use. Preliminary results show no device-related complications and a satisfactory 6-month angiographic patency.
Medical subject headings
- Aorta
- Coronary Artery Bypass
- Saphenous Vein