Enhanced notification of critical ventilator events.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 16049226.
- Also identified by PMC identifier 1294028.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Mechanical ventilators are designed to generate alarms when patients become disconnected or experience other critical ventilator events. However, these alarms can blend in with other accustomed sounds of the intensive care unit. Ventilator alarms that go unnoticed for extended periods of time often result in permanent patient harm or death. We developed a system to monitor critical ventilator events through our existing hospital network. Whenever an event is identified, the new system takes control of every computer in the patient's intensive care unit and generates an enhanced audio and visual alert indicating that there is a critical ventilator event and identifies the room number. Once the alert is acknowledged or the event is corrected, all the computers are restored back to the pre-alert status and/or application. This paper describes the development and implementation of this system and reports the initial results, user acceptance, and the increase in valuable information and patient safety.
Medical subject headings
- Hospital Information Systems
- Monitoring, Physiologic
- Ventilators, Mechanical