Cup alignment error model for total hip arthroplasty.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 16056040.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Almost all computer-assisted orthopaedic surgery systems that rely on the anterior pelvic plane definition, such as in computed tomography and magnetic resonance image-based, fluoroscopy-based, and nonimage total hip replacement approaches, are derived from identifying two pairs of pelvic bony landmarks: anterior superior iliac spines and the pubic tubercles. Although these systems strive to achieve cup alignment accuracy of approximately 1 degree, even a minor failure to correctly identify these anatomic landmarks can lead to higher inaccuracies in the final cup alignment. This study shows how to examine the effects of these inaccuracies on the final acetabular cup implant orientation during total hip replacement by generating a kinematic model, which then is simulated. Simulation results indicate that, for example, a total error of 4 mm in measuring the anterior superior iliac spine and the pubic tubercles would result in a final cup orientation of 47 degrees and 27 degrees in abduction and version respectively, resulting in a 2 degrees abduction error and 7 degrees error in version when targeting 45 degrees abduction and 20 degrees version results. These calculations can be repeated for any error values.
Medical subject headings
- Acetabulum
- Arthroplasty, Replacement, Hip
- Computer Simulation
- Hip Joint
- Medical Errors
- Models, Anatomic