Non-invasive measurement of intraocular pressure by near-infrared spectroscopy.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 16086952.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To assess a new non-contact method for non-invasive measurement of intraocular pressure (IOP) using near-infrared spectroscopy. Experimental study. Eleven porcine eyes were cannulated into the anterior chamber by a needle connected with a water column. Measuring the near-infrared spectrum in diffuse reflectance out of the anterior chamber by near-infrared spectroscopy, IOP was determined and compared with the height of the water column which was varied between 0 cm and 100 cm. In total, 363 near-infrared spectrometric measurement values were obtained. Correlating the near-infrared spectrometric measurements with the height of the water column showed a relationship with a correlation coefficient of r(2) = .858 and a steepness of the regression line of .96. The results of the present pilot study suggest that near-infrared spectrometric tonometry may present a new technology for non-invasive and non-contact IOP measurement.
Medical subject headings
- Intraocular Pressure
- Spectroscopy, Near-Infrared
- Tonometry, Ocular