Stereopsis-guided brain shift compensation.

Sun, Hai; Lunn, Karen E; Farid, Hany; Wu, Ziji; Roberts, David W; Hartov, Alex; Paulsen, Keith D · IEEE Trans Med Imaging · 2005

case_report · Level V

Where this comes from

Abstract

Brain deformation models have proven to be a powerful tool in compensating for soft tissue deformation during image-guided neurosurgery. The accuracy of these models can be improved by incorporating intraoperative measurements of brain motion. We have designed and implemented a passive intraoperative stereo vision system capable of estimating the three-dimensional shape of the surgical scene in near real-time. This intraoperative shape is compared with the cortical surface in the co-registered preoperative magnetic resonance (MR) volume for the estimation of the cortical motion resulting from the open cranial surgery. The estimated cortical motion is then used to guide a full brain model, which updates a preoperative MR volume. We have found that the stereo vision system is accurate to within approximately 1 mm. Based on data from two representative clinical cases, we show that stereopsis guidance improves the accuracy of brain shift compensation both at and below the cortical surface.

Medical subject headings