To do or not to do a preinduction check-up of the anesthesia machine.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 16115990.
- Also identified by DOI 10.1213/01.ANE.0000175210.13435.82.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The anesthesia machine check is an integral part of the anesthesiologist's daily routine. It is standard operating procedure to check the high- and low-pressure systems of the machine as well as other integral parts that are accessible. Many new anesthesia machines offer a self-testing capability, but older ones with fewer electronics on board are still widely used. Whether the machines self test or not, each machine contains a CO2 absorber and a circuit that may be prone to problems. In our case we encountered an open CO2 absorber after a service of the machine had been performed between the cases without our knowledge. We were unable to ventilate the patient during induction. The presence of a backup self-filling manual ventilation bag was invaluable in preventing an otherwise inevitable emergency.