In vivo wear and migration of highly cross-linked polyethylene cups a radiostereometry analysis study.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 16124954.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In 50 cemented hip arthroplasties, wear and migration of the polyethylene (PE) cups were measured with radiostereometric analysis for a period of 2 years. Twenty had a normal gamma-in-air-sterilized PE, another 20 had a PE sterilized with 30000 Gy followed by heat stabilization (Duration; Stryker Orthopaedics, Mahwah, NJ), and 10 had highly cross-linked PE cups irradiated with 100000 Gy (Crossfire; Stryker Orthopaedics). In the initial 2 months, head penetration (creep) was 63 microm on average for the 3 groups. From 2 to 24 months, the mean proximal head penetration (wear) was 156 microm for standard PE, 138 microm for stabilized PE (P = .45), and 23 microm for highly cross-linked PE (P < .001; analysis of variance). The low in vivo wear rate for highly cross-linked cups was not at the expense of higher migration or less favorable clinical outcome and looks promising.
Medical subject headings
- Hip Prosthesis
- Polyethylenes
Anatomy
- hip