Pulse oximetry in severe anaemia.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 1613193.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Measurement of arterial oxygen saturation by pulse oximetry was performed in two patients with acute and chronic anaemia (haemoglobin concentrations: 2.9 mmol/l (4.7 g/dl) and 1.9 mmol/l (3.0 g/dl), respectively) using a Radiometer OXI and a Nellcor N-200 pulse oximeter. The two oximeters read alternating different values in the two patients. In conclusion, pulse oximeters are able to give a value for oxygen saturation even at extreme anaemia, and when a high value is given, it possibly reflects arterial oxygen saturation. The value of pulse oximetry in severe anaemia is discussed.
Medical subject headings
- Anemia
- Hemoglobins