A study of the effect of tibial tray rotation on a specific mobile bearing total knee arthroplasty.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 16139718.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We presumed that a particular mobile-bearing knee prosthesis would cope with tibial tray rotation about its longitudinal axis. We studied the tibiofemoral and patellofemoral joints to assess this. The prosthesis was mounted on a specially designed jig. The tray was externally rotated in 5 degrees increments and put through a range of flexion. We found that increasing tibial tray rotation caused excessive polyethylene loading and, eventually, medial femoral component liftoff. The patellofemoral joint showed increasing lateral patella facet contact. Surprisingly, this mobile-bearing prosthesis could not cope with tibial tray rotation. It required the tibial tray to be neutrally aligned to the femoral component to avoid excessive polyethylene loading.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Prosthesis
Anatomy
- knee
- tibia