An inverse methodology for high-frequency RF coil design for MRI with de-emphasized B1 fields.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 16189971.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
An inverse methodology for the design of biologically loaded radio-frequency (RF) coils for magnetic resonance imaging applications is described. Free space time-harmonic electromagnetic Green's functions and de-emphasized B1 target fields are used to calculate the current density on the coil cylinder. In theory, with the B1 field de-emphasized in the middle of the RF transverse plane, the calculated current distribution can generate an internal magnetic field that can reduce the central overemphasis effect caused by field/tissue interactions at high frequencies. The current distribution of a head coil operating at 4 T (170 MHz) is calculated using an inverse methodology with de-emphasized B1 target fields. An in-house finite-difference time-domain routine is employed to evaluate B1 field and signal intensity inside a homogenous cylindrical phantom and then a complete human head model. A comparison with a conventional RF birdcage coil is carried out and demonstrates that this method can help in decreasing the normal bright region caused by field/tissue interactions in head images at 170 MHz and higher field strengths.
Medical subject headings
- Brain
- Electromagnetic Phenomena
- Magnetic Resonance Imaging
- Models, Biological
- Radio Waves
- Transducers