Difficult ventilation with a double-lumen endotracheal tube: an unusual manufacturing defect.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 16192527.
- Also identified by DOI 10.1213/01.ane.0000167637.36187.22.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report an unexpected difficult ventilation with a double-lumen endotracheal tube in a patient receiving left upper lobe lung mass resection. The manufacturing defect in both limbs of the Opti-Port Right Angle Double Swivel Connector of the double-lumen tube resulted in this problem. This defect is difficult to localize by the usual recommended methods. We discuss a modified algorithm for difficult ventilation with a double-lumen endotracheal tube. Difficult ventilation occurred during general anesthesia as the result of a manufacturing defect in both limbs of the connector of the double-lumen endotracheal tube. The problem was resolved with a careful approach, and there were no serious consequences.