Difficult ventilation with a double-lumen endotracheal tube: an unusual manufacturing defect.

Chen, Hung-Shu; Jawan, Bruno; Tseng, Chia-Chih; Cheng, Kwok-Wai; Wang, Chih-Hsien · Anesth Analg · 2005

case_report · Level V

Where this comes from

Abstract

We report an unexpected difficult ventilation with a double-lumen endotracheal tube in a patient receiving left upper lobe lung mass resection. The manufacturing defect in both limbs of the Opti-Port Right Angle Double Swivel Connector of the double-lumen tube resulted in this problem. This defect is difficult to localize by the usual recommended methods. We discuss a modified algorithm for difficult ventilation with a double-lumen endotracheal tube. Difficult ventilation occurred during general anesthesia as the result of a manufacturing defect in both limbs of the connector of the double-lumen endotracheal tube. The problem was resolved with a careful approach, and there were no serious consequences.