Frequency effect in fretting wear of Co-28Cr-6Mo versus Ti-6Al-4V implant alloys.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 16211570.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A piezo-electrically driven fretting testing device has been constructed and fretting release and release rates have been determined with highest accuracy, using a radiotracer technique. First results on the fretting release and release rate of titanium alloy fretting pads against cobalt-chrome alloy fatigue specimens are reported. The frequency dependency of fretting release has been determined between 1 and 8 Hz and shows higher release rates for low frequencies, thus indicating that accelerated testing of materials and components of artificial joints must be analyzed extremely carefully. The present experiments under simple conditions present a base-line study for step-wise applying more complex and realistic testing conditions and for using radiotracer methods to quantify fretting release in simulated testing of artificial hip- and knee-prostheses.
Medical subject headings
- Joint Prosthesis
- Materials Testing
- Titanium
- Vitallium