Cracking and impingement in ultra-high-molecular-weight polyethylene acetabular liners.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 16214008.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to determine the prevalence of crack formation in conventional ultra-high-molecular-weight polyethylene cups and its association with rim impingement, oxidation, and time in situ. One hundred twenty acetabular cups were retrieved during revision total hip arthroplasty. In 40% (48/120) of the retrieved specimens, multiple subsurface cracks of several millimeters in length were revealed by transillumination. In 5 specimens, full thickness cracks led to fragmentation of the liner before revision. Thirty-eight (32%) liners had regions of moderate to severe impingement damage to the rim; cracks were initiated at the site of impingement in all but 1 liner (P < .0001). Cracks commonly occur in conventional ultra-high-molecular-weight polyethylene liners, often after neck impingement and almost always in association with oxidation of the polymer.
Medical subject headings
- Hip Prosthesis
- Polyethylenes
Anatomy
- hip