A Monte Carlo study of intersource effects in dome-type applicators loaded with LDR Cs-137 sources.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 16226817.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In this study, the dose rate distributions produced by low dose rate Cs-137 sources loaded in afterloadable dome applicators are studied using the Monte Carlo method. Dose differences between Monte Carlo results and calculations done using the superposition principle are within 1-3% in front of the applicator and between 3 and 10% near and along the longitudinal source axis. Consequently, the real doses to lateral vaginal wall, rectum and bladder are very close to the doses estimated applying the superposition principle, while the dose to the vaginal cuff has been overestimated by up to 10%.
Medical subject headings
- Brachytherapy
- Monte Carlo Method
- Radiotherapy Planning, Computer-Assisted