In vivo comparative wear study of traditional and highly cross-linked polyethylene in total hip arthroplasty.
Where this comes from
- Record sourced from PubMed, PMID 16230239.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In this study, we compare the in vivo wear performance of electron beam-irradiated, postirradiation-melted, highly cross-linked polyethylene (HXLPE) and traditional UHMWPE via the Martell method. Seventy hips with HXLPE performed at the Massachusetts General Hospital had 138 radiograph pairs for wear analysis and a 31.2-month average follow-up (range, 24-44 months). An age-matched, sex-matched, and body mass index-matched subgroup of 111 hips with 214 acceptable radiograph pairs and a 4-year follow-up from our previously published study on traditional polyethylene performed at Rush-Presbyterian-St. Luke's Medical Center served as a control group. Martell wear analysis was performed for each group. Overall and steady-state wear rates were compared via a specialized t test. The steady-state wear in the HXLPE arm was observed after 2.0 years, was 0.007 mm/y, and was significantly less than the steady-state wear in the traditional arm (0.174 mm/y) (P = .003). Highly cross-linked polyethylene penetration rate was not affected by sex, age, activity, or body mass index by Mann-Whitney analysis.
Medical subject headings
- Hip Prosthesis
- Polyethylene
- Prosthesis Failure
Anatomy
- hip