Is the valve OK or not? Immediate evaluation of a replaced aortic valve.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 16243981.
- Also identified by DOI 10.1213/01.ANE.0000181339.39448.F0.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Transesophageal echocardiography is a crucial tool in intraoperative evaluation of newly implanted/repaired heart valves because suspected valvular malfunction needs to be identified and sometimes surgically corrected. Although color Doppler is often adequate in evaluating the expected regurgitant jets, as well as excluding pathologic paravalvular leaks, spectral Doppler techniques are the most commonly used methods for estimating transvalvular gradients in the operating room. However, these methods are subject to a variety of confounding factors, including subvalvular gradients and pressure recovery. Other methods of valve area estimation should also be used when evaluating a prostethic aortic valve, including the continuity equation and the left ventricular outflow tract/aortic valve velocity ratio.
Medical subject headings
- Aortic Valve
- Aortic Valve Stenosis
- Heart Valve Prosthesis